Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: AVO DIGIMINOR 2000: (1 year) CAL VER /5500 DATE: 27-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 96 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Specifications # # Compatibility: # 5500/CAL or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP {Procedure not formally tested with UUT.} 1.001 DISP {See procedure README file.} 1.002 ASK- N P J F W 1.003 ASK+ K 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {RESISTANCE TESTS} 1.007 DISP Connect the 5500A and the UUT as follows: 1.007 DISP [32] 5500A AUX Hi to UUT + 1.007 DISP [32] 5500A AUX LO to UUT - 1.007 DISP [32] 5500A NORMAL Hi to UUT + 1.007 DISP [32] 5500A NORMAL LO to UUT - 1.008 HEAD RESISTANCE TESTS: {2000 Ohm Range} 1.009 5500 2000 1000Z 1.0% 1U CW 2.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 2.002 5500 200 100.0kZ 1.0% 0.1U CW 3.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 3.002 DISP Connect the 5500A and the UUT as follows: 3.002 DISP [32] 5500A NORMAL Hi to UUT + 3.002 DISP [32] 5500A NORMAL LO to UUT - 3.003 5500 20 10.00MZ 2.0% 0.01U 2W 4.001 HEAD {DC VOLTAGE TESTS} 4.002 HEAD DC VOLTAGE TESTS: {2V Range} 4.003 5500 2 1.900V 0.25% 0.001U 2W 5.001 HEAD DC VOLTAGE TESTS: {20V Range} 5.002 5500 20 19.00V 0.25% 0.01U 2W 6.001 5500 20 -19.00V 0.25% 0.01U 2W 7.001 HEAD DC VOLTAGE TESTS: {200V Range} 7.002 5500 200 190.0V 0.25% 0.1U 2W 8.001 HEAD DC VOLTAGE TESTS: {1000V Range} 8.002 5500 1000 1000V 0.25% 1U 2W 9.001 HEAD {AC VOLTAGE TESTS} 9.002 HEAD AC VOLTAGE TESTS: {2V Range} 9.003 5500 2 1.900V 1.0% 0.003U 60H SI 2W 10.001 5500 2 1.900V 1.0% 0.003U 2kH SI 2W 11.001 HEAD AC VOLTAGE TESTS: {20V Range} 11.002 5500 20 19.00V 1.0% 0.03U 60H SI 2W 12.001 5500 20 19.00V 1.0% 0.03U 2kH SI 2W 13.001 HEAD AC VOLTAGE TESTS: {200V Range} 13.002 5500 200 190.0V 1.0% 0.3U 60H SI 2W 14.001 5500 200 190.0V 1.0% 0.3U 1kH SI 2W 15.001 HEAD AC VOLTAGE TESTS: {1000V Range} 15.002 5500 1000 750V 1.0% 3U 60H SI 2W 16.001 5500 1000 750V 1.0% 3U 400H SI 2W 17.001 HEAD {DC CURRENT TESTS} 17.002 DISP Connect the 5500A and the UUT as follows: 17.002 DISP [32] 5500A AUX Hi to UUT A 17.002 DISP [32] 5500A AUX LO to UUT - 17.003 HEAD DC CURRENT TESTS: {20mA Range} 17.004 5500 20 19.00mA 1.0% 0.01U 2W 18.001 HEAD DC CURRENT TESTS: {2A Range} 18.002 5500 2 1.900A 1.0% 0.001U 2W 19.001 HEAD {AC CURRENT TESTS} 19.002 HEAD AC CURRENT TESTS: {20mA Range} 19.003 5500 20 19.00mA 2.0% 0.03U 60H SI 2W 20.001 5500 20 19.00mA 2.0% 0.03U 1kH SI 2W 21.001 HEAD AC CURRENT TESTS: {2A Range} 21.002 5500 2 1.900A 2.0% 0.003U 60H SI 2W 22.001 5500 2 1.900A 2.0% 0.003U 1kH SI 2W 23.001 END